Ametek DCS-E 1kW Series Programming Manual Instrukcja Użytkownika Strona 19

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 32
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 18
DLM-E Series and DCS-E Series Isolated Analog Programmer Operation
2.3 Remote Programming Configuration
2.3.1 Voltage Source Programming of Output Voltage
SW1 Switch Settings
SW1-1 ON (1)= 0–5V
SW1-1 OFF (0)= 0–10V
291213 1
25
14
Set up for voltage source programming of the output voltage as follows:
1. Set SW1-1, V PROG, to ON (1) for 0-5VDC programming range.
2. Set SW1-1, V PROG, to OFF (0) for 0-10VDC programming range.
3. Connect the external programming voltage source to the ISOLATED ANALOG
PROGRAMMER connector, J4, with positive to Pin-9 and negative to Pin-12.
4. Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG PROGRAMMER
connector, J4, to Pin-2 to enable remote control.
5. Program the other parameters to the desired limit values.
M51A Option 2-7
Przeglądanie stron 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 31 32

Komentarze do niniejszej Instrukcji

Brak uwag