Controller Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 37
Table 2-2. SCPI Error Codes
0
The error queue is empty.
2
When a command is issued to an empty slot
11
Command is not correct for the specific module type.
12
Specific command may not get implemented for the module type
14
Command will be implemented in the future
15
System is busy executing a previous command, the command causing this error will
not be executed.
206
TrigChannel not available
Requested trigger channel is already used in the system
251
Group creation failed for stated reason.
252
Module inhibited asserted
Module Inhibit is asserted in the RFPC for the ReFlex Power™ system.
253
The LIST being created, is too long.
-102
An unrecognized command or data type was encountered.
-200
A command could not be executed due to the current condition of the module.
-222
Parameter exceeded range of valid values.
-292
Requested File name either not found or the name was not valid.
-293
Requested name already exists.
-331
A valid command could not be executed due to some internal/earlier error.
-350
The error queue can contain up to ten entries. If more than ten error/event conditions
are logged before a SYSTem:ERRor? Query, an overflow will occur; the last queue
entry will be overwritten with error -350. When the queue overflows, the least recent
error/events remain in the queue and the most recent error/events are discarded.
2.8
LXI™ AND SCPI CONFORMANCE INFORMATION
The ReFlex Power™ Controller Ethernet is IEEE 802.3 and LXI™ (LAN eXtensions
for Instrumentation) class C compliant.
The syntax of all SCPI commands implemented by the ReFlex Power™ system
and documented in this manual, are either SCPI commands confirmed in the SCPI
1999.0 Specification, Volume 2: Command Reference, or they are customized
commands, not part of the SCPI definition, but do conform to SCPI syntax.