AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
148 Manual P/N 7003-961 Rev AA
6.4 Programming Output Transients
Output transients are used to:
Synchronize output changes with a particular phase of the voltage waveform.
Synchronize output changes with internal or external trigger signals.
Simulate surge, sag, and dropout conditions with precise control of duration and phase.
Create complex, multi-level sequences of output changes.
Create output changes that have rapid or precise timing requirements.
The following AC/DC source functions are subject to transient control:
AC output voltage
DC output voltage
Frequency
Start phase angle
AC/DC voltage slew rate
Frequency slew rate
The following transient modes can be generated:
Step generates a single triggered output change.
Pulse generates an output change which returns to its original state after some
time period.
List generates a sequence of output changes, each with an associated dwell
time or paced by triggers.
Fixed turns off the transient functions, which means that only the IMMediate
values are used as the data source for a particular function.
After a *RST or Device Clear command, all functions are set to FIXed, which turns off the
transient functions.