AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
98 Manual P/N 7003-961 Rev AA
4.18 Source Subsystem - Mode
The MODE command allows switching between the different output modes available on the
AC/DC power source. The MX Series with a –1 or –3 provides either AC or DC mode while
the MX/RS Series with a –1Pi or –3Pi adds AC+DC mode as well. When switching modes,
the output is automatically set to zero to prevent hot switching of the output. After a mode
command, the output voltage needs to be programmed to the desired setting.
Subsystem Syntax
[SOURce:]
MODE AC | DC | ACDC Sets the output mode
MODE
The mode command switches the output voltage between the available output modes.
Command Syntax [SOURce:]MODE
Parameters AC | DC | ACDC
Example MODE AC
Query Syntax [SOURce:]MODE?
Returned Parameters <CRD>
Related Commands PONS:VOLT:MODE