6.6.10 Selecting the Measurement Trigger Source
The trigger system is waiting for a trigger signal in the Initiated state. Before you generate a
trigger, you must select a trigger source. To select the SYNC Trigger to start the
measurement at specific phase angles in the waveform:
TRIGger:ACQuire:SOURce SYNC
TRIGger:SYNC:SOUR PHAS
TRIGger:SYNC:PHASE 0
To select IEEE-488 bus triggers (group execute trigger, device trigger, or *TRG command),
use:
TRIGger:ACQuire:SOURce BUS
To select the signal driving the Trigger Out from the transient, use:
TRIGger:ACQuire:SOURce TTLTrg
6.6.11
Generating Measurement Triggers
Providing that you have specified the appropriate trigger source, you can generate triggers
as follows:
By sending one of the following over the IEEE-488:
TRIGger:ACQuire:IMMediate
*TRG
a group execute trigger (GET)
By waiting for a specific phase angle of the waveform
By generating an output transient that causes the Trig Out to output a pulse.
By pressing the front panel START soft key in the HARMONICS/WAVEFORM when the
unit is operating in local mode.