AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
184 Manual P/N 7003-961 Rev AA
IEC413:SWEep:CURRent:RESonant?
This command will return the measured harmonic resonant current. This data is available
after the sweep is completed. If no resonant current is found, the command will return a null
char.
Command Syntax IEC413:SWEep:CURRent:RESonant?
Parameters none
Query Syntax IEC413:SWE:CURR:RES?
Returned Parameters <NR2>,.... ,<NRn>
IEC413:SWEep:FREQuency:RESonant?
This command will return the frequency resonant data points that correspond to the
harmonic current data points. This data is available after the sweep is completed. If no
resonant current is found, the command will return a null char.
Command Syntax IEC413:SWEep:CURRent:RESonant?
Parameters none
Query Syntax IEC413:SWE:FREQ:RES?
Returned Parameters <NR2>,.... ,<NRn>
IEC413:HARMonics:LEVel
This command will set the level for the harmonics test. The command will accept either one
parameter or a sequence of parameters from the second harmonic to the fortieth harmonic.
In case of one parameter, the level will be set for the selected harmonic number specified by
IEC413:INDV:HARM:NUMBer. The IEC413:STATe must be in the off position and the user
class must be selected to accept the command.
Command Syntax IEC413:HARMonics:LEVel <Nrf+> [,<Nrf>,....<Nrfn>]
Parameters 0 to 20%
Examples IEC413:HARM:LEV 4,5.......,8
Query Syntax IEC413:HARM:LEV?
Returned Parameters <NR2>,...,<NR2>
IEC413:HARMonics:PHASe
This command will set the phase angle of the harmonics wave form relative to the
fundamental. The command will accept either one parameter or a sequence of parameters
from the second harmonic to the fortieth harmonic. In case of one parameter, the phase will
be set for the selected harmonic number specified by IEC413:INDV:HARM:NUMB. The
IEC413:STATe must be in the off position and the user class must be selected to accept the
command.
Command Syntax IEC413:HARMonics:PHASe <Nrf+> [,<Nrf>,....<Nrfn>]
Parameters 0 to 360
Examples IEC413:HARM:PHAS 180,270,.......,90
Query Syntax IEC413:HARM:PHAS?
Returned Parameters <NR2>,...,<NR2>