Ametek BPS Series Programming Manual Instrukcja Użytkownika Strona 44

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 226
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 43
AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
44 Manual P/N 7003-961 Rev AA
4.2.4 Voltage Measurement
CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale] <NRf>
This command initiates the calibration of the rms voltage measurement at full scale and at
ambient temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale]
Parameters <NRf> (actual rms output voltage measured with external device)
Examples CAL:MEAS:VOLT 120
Query Syntax CALibrate:MEASure:VOLTage?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:[:AC][:FSCale]
CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale] <NRf>
This command initiates the calibration of the DC voltage measurement at full scale and at
ambient temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale]
Parameters <NRf> (actual DC output voltage measured with external device)
Examples CAL:MEAS:VOLT:DC 120
Query Syntax CALibrate:MEASure:VOLTage:DC?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]
CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO <NRf>
This command initiates the offset adjustment of the DC voltage measurement at ambient
temperature.
Command Syntax CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO
Parameters <NRf> (0 or desired offset value)
Examples CAL:MEAS:VOLT:DC:ZERO 0
Query Syntax CALibrate:MEASure:VOLT:DC:ZERO?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO
CALibrate:MEASure:VOLTage:TEMPerature[:AC][:FSCale] <NRf>
This command initiates the calibration of the rms voltage measurement at full scale and at
elevated temperature.
Command Syntax CALibrate:MEASure:VOLTage:TEMPerature[:AC][:FSCale]
Parameters <NRf> (actual rms output voltage measured with external device)
Examples CAL:MEAS:VOLT:TEMP 120
Query Syntax CALibrate:MEASure:VOLTage:TEMPerature?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage[:AMBient][:AC][:FSCale]