Programming Manual Lx \ Ls Series II
98
4.23 Status Subsystem Commands
This subsystem programs the Lx\Ls Series status registers. There are four groups of status
registers; Operation, Questionable, Questionable Instrument ISummary and Standard Event.
The Standard Event group is programmed with Common commands. The Operation,
Questionable, and Instrument ISummary status groups each consist of the following five registers:
Refer to Chapter 7 for more information about the status registers.
Subsystem Syntax
STATus
:PRESet Presets all enable and transition registers to power-on
:OPERation
[:EVENt] Returns the value of the event register
:CONDition Returns the value of the condition register
:ENABle <n> Enables specific bits in the Event register
:NTRansition<n> Sets the Negative transition filter
:PTRansition<n> Sets the Positive transition filter
:QUEStionable
[:EVENt] Returns the value of the event register
:CONDition Returns the value of the condition register
:ENABle <n> Enables specific bits in the Event register
:NTRansition<n> Sets the Negative transition filter
:PTRansition<n> Sets the Positive transition filter
:INSTrument
:ISUMmary
[:EVENt] Returns the selected phase's event register value
:CONDition Returns the selected phase's condition register value
:ENABle <n> Enables specific bits in the selected phase's Event
register
:NTRansition<n> Sets the selected phase's Negative transition filter
:PTRansition<n> Sets the selected Phase's Positive transition filter
STATus:PRESet
This command sets the Enable, PTR, and NTR registers of the status groups to their power-on
values. These values are:
Enable Registers: all bits set to 0 (OFF)
PTR Registers: all defined bits set to 1 (ON)
NTR Registers: all bits set to 0 (OFF)
Command Syntax STATus:PRESet
Parameters None
Examples STAT:PRES
CAL Interface is computing new calibration constants