8.7.3 MS704 Steady State Commands
Steady state tests can be started by sending one of the MS704:NORM commands. Available
normal state commands are listed below.
MS704:NORMal:SSLimits[:ALL]
This command will run all steady state normal operation mode tests.
Command Syntax MS704:NORMal:SSLimits[:ALL]
Parameters none
Examples MS704:NORM:SSL
Query Syntax none
MS704:NORMal:SSLimits:NOMinal
This command will run the steady state normal operation nominal voltage test.
Command Syntax MS704:NORMal:SSLimits:NOMinal
Parameters none
Examples MS704:NORM:SSL:NOM
Query Syntax none
MS704:NORMal:SSLimits:LOW
This command will run the steady state normal operation low voltage test.
Command Syntax MS704:NORMal:SSLimits:LOW
Parameters none
Examples MS704:NORM:SSL:LOW
Query Syntax none
MS704:NORMal:SSLimits:HIGH
This command will run the steady state normal operation high voltage test.
Command Syntax MS704:NORMal:SSLimits:HIGH
Parameters none
Examples MS704:NORM:SSL:HIGH
Query Syntax none
MS704:NORMal:SSLimits:UNBalance
This command will run the steady state normal operation voltage unbalance test. This test is only
available in the 3 phase mode.
Command Syntax MS704:NORMal:SSLimits:UNBalance
Parameters none
Examples MS704:NORM:SSL:UNB
Query Syntax none