OUTPut:DFI:SOURce
This command selects the source for DFI events. The choices are:
QUEStionable Questionable summary bit
OPERation Operation summary bit
ESB Standard Event summary bit
RQS Request Service summary bit
OFF Never true
Command Syntax OUTP:DFI:SOUR<source>
Parameters QUES | OPER | ESP | RQS | OFF
*RST Value OFF
Examples OUTP:DFI:SOUR OPER
Query Syntax OUTPut:DFI:SOUR?
Returned Parameters <CRD>
Related Commands OUTP:DFI
OUTPut:PON[:STATe]
This command selects the power-on state of the AC source. The following states can be selected:
RST Sets the power-on state to *RST. Refer to the *RST command as
described later in this chapter for more information.
RCL0 Sets the power-on state to *RCL 0. Refer to the *RCL command as
described later in this chapter for more information.
Command Syntax OUTPut:PON[:STATE] <state>
Parameters RST | RCL0
Examples OUTP:PON:STAT RST
Query Syntax OUTPut:PON:STATe?
Returned Parameters <CRD>
Related Commands *RST *RCL
OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an overvoltage (OV), overcurrent
(OC), overtemperature (OT), or remote inhibit (RI) fault condition is detected. All conditions that
generated the fault must be removed before the latch can be cleared. The output is then restored
to the state it was in before the fault condition occurred.
Command Syntax OUTPut:PROTection:CLEar
Parameters None
Examples OUTP:PROT:CLE
Related Commands OUTP:PROT:DEL *RCL *SAV