Programming Manual Lx \ Ls Series II
187
8.3.3 Language Extension
The following additional capabilities not part of the ABLE syntax - are supported:
All IEEE488.2 (Dot2) commands can be accessed without having to switch to SCPI
mode.
VOLTS, CURL, FREQ, RNG, RNGF, CLS and OPN can be queried by appending a
question mark. The return data reflects the active setting.
8.3.4 Serial Poll Status Bytes
STATUS
DESCRIPTION
64
Servo Error. This is caused by an open servo or a servo that could not
maintain the output voltage to the set value.
67
Over voltage error. Output voltage exceeds program value.
71
A phase current limit error.
72
B phase current limit error.
73
C phase current limit error.
74
Syntax error. Command not recognized by the source.
75
Command error. Data out of range for programmed parameter.
76
Buffer over flow. Input message exceeds the input buffer size.
78
This error message will not be generated.
79
Measurements complete. TEST n and TST commands will generate
status byte 79 when complete.
Table 8-9: ABLE languange - Serial Poll Status Bytes.