Programming Manual Lx \ Ls Series II
92
VOLTage:ALC:SOURce
These commands select the source from which the output voltage is sensed. The following
voltage sense sources can be selected:
INTernal This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.
EXTernal This senses the output voltage at the user's sense terminals, which
allows remote voltage sensing at the load.
Command Syntax [SOURce:]VOLTage:ALC:SOURce<source>
Parameters INTernal | EXTernal
*RST Value INTernal
Examples VOLT:ALC:SOUR EXT
Query Syntax [SOURce:]VOLTage:ALC:SOURce?
Returned Parameters <CRD>
Related Commands VOLT:SENS:DET
Note: The VOLT:ALC:SOUR command is an alias for the VOLT:SENS:SOUR command. Both
perform the same function. This is done for backward compatability with the Agilent 6834B.
VOLTage
Phase Selectable
This command programs the ac rms output voltage level of the AC source. The maximum peak
voltage that the AC source can output is 425 V peak. This includes any combination of voltage
and function shape values. Therefore, the maximum value that can be programmed depends on
the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum voltage that can be
programmed is 300 V rms.
Note: You cannot program a voltage that produces a higher volt-second on the output than a
300V rms sinewave.
Command Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]<NRf+>
Parameters 0 to 300 (for sinewaves)
Unit V (rms voltage)
*RST Value 1 volt
Examples VOLT 250 VOLT:LEV 25
Query Syntax [SOURce:]VOLTage[:LEVel]
[:IMMediate][:AMPLitude]?
Returned Parameters <NR3>
Related Commands VOLT:MODE VOLT:TRIG FUNC:SHAP
Przeglądanie stron 95