Programming Manual - Rev S Lx \ Ls Series II
230
"Trigger received before
requested number of
pre-trigger readings"
Data acquisition pre-
trigger buffer not
filled yet.
Hold off trigger or reduce pre-trigger
delay.
"Requested RMS current
too high for voltage
range"
Max RMS current is
function of voltage
range selected.
Reduce programmed RMS current
limit or select low voltage range.
"Waveform data not
defined"
No waveform name
specified
Specify waveform name before
sending waveform data.
"VOLT,VOLT:SLEW, and
FUNC:SHAPe modes
incompatible"
Conflict between
wave shape and
programmed slew
Reduce slew or change waveform
type.
Measurement data
out of range.
Too much data in
output buffer.
Check receive mode on application
program. Program is not reading data
sent by AC source.
"Command cannot be
given with present
SYST:CONF setting"
Command conflicts
with available
hardware or firmware
option settings.
Check configuration for available
options and features.
- Output voltage does
not match program
value when ALC is
on.
- Over load
- Voltage kick-back
- No output voltage
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting.
Output voltage is driven above
programmed voltage by external
influence (Load, voltage kickback,
etc.)
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting
Amplifier heat sink
temp. too high.
Reduce load. Ensure proper air flow
and exhaust clearance. Check fan(s)
for operation.
Could not sync to
external sync signal.
External sync signal missing,
disconnected or out of range.
Initial settings could
not be recalled at
power-up.
Save power on settings again to
overwrite old content.
Hardware
configuration settings
could not be recalled
at power-up.
Contact AMETEK service department
at service@programmablepower.com
to obtain instructions on restoring
configuration data.
Memory corrupted
during power-up.
"Calibration memory lost"
Calibration data lost
during power-up.
Contact AMETEK service department
at service@programmablepower.com
to obtain instructions on restoring
calibration data or recalibrate unit.
One or more
transient list
parameters missing.